title graphic AIF graphic

Hazebuster Deconvolution for OSX by Vaytek

 

When first setting up the .mic files you will see an option for signal to noise settings (0-30 inclusive). After contacting Vaytek we suggest that all users use a setting of 1. If you would like to wholly understand why we're using this setting please ask one of the AIF LM staff.

Be sure to convert all files to 12 bit (dynamic range 0-4095) before starting deconvolution and set the rapid deconvolution option to 12 bit.

*You may want to convert the final 12 bit image back to 16 bit by multiplying the pixels by 16.

Download PDF version of the VATEK instructions.*right click <save as> [To save using a PC]

(These are included further down on this page as JPEG files)

This deconvolution is the quickest, most used and least robust deconvolution software available at the AIF. 

Deconvolution is used to make a wide field epifluorescence (or any other) image and mathematically performs a sort of "sharpen" function with a resulting image that looks as though it was collected on a confocal microscope.  [More on deconvolution here.]

Instructions 

Parameters

Objective lens power 10X 20X 40X air 40X oil 60X 60X with 1.5X eyepiece
N.A. 0.30 0.40 0.60 1.00 1.40 1.40
Pixel size mm 0.66 0.330 0.165 0.165 0.110 0.073

The values for pixel size are for 1 X 1 binning.  If you use 2 X 2 binning, simply multiply by 2.  For 3 X 3 binning, multiply by 3.
*Test question:  For 4 X 4 binning, what would you do?

Objective lens 1.25X to 40X air 25X to 100X oil
Refractive Index 1.00 1.515

 

fluorophore DAPI Low blue FITC Rhodamine Cy5
wavelength 0.460 0.460 0.520 0.600 0.685

 






 return to index next page
Page revised and updated 01/07/2005 by D.A.G.