title graphic AIF graphic

whole_microscope_arrowed.jpg (170588 bytes)Background:

Digital Station #6 was purchased with funds on a Shared Instrumentation Grant to meet the needs of multicolor FISH and other multiprobe high resolution light microscope applications.

Instrumentation:

This is a dual use instrument.  The microscope may be driven by the SKY software or by IPLab.  The information on this web page is for IPLab use only.

For methods for a paper, you might use the following specifications:
  Olympus BX61
  100X N.A. 1.4 UPlanSApo
  Cooke Sensicam QE cooled CCD camera
  IPLab 4 software

Instructions:

These instructions are for using the system with the Cooke Sensicam CCD camera only.

Jump to:


ALL OF THESE INSTRUCTIONS MUST BE FOLLOWED FOR THE SYSTEM TO OPERATE.
TREAT THIS AS A CHECKLIST.  PRETEND YOU ARE CHECKING OFF FOR THE LAUNCH OF YOUR ROCKET SHIP.
YOU DON'T WANT TO CRASH OR EXPLODE.

1.  Mercury arc lamp.

hg-off.gif (13791 bytes)
Mercury lamp off
      hg-on.gif (13685 bytes)
Mercury lamp on

2.  The correct filter turret must be installed in the microscope.  This is the correct turret.
8positionwheel.jpg (12510 bytes)
Note that there are 8 positions. 
If this is not the wheel installed in the microscope, you must switch to the other wheel.   Click here for instructions how to do this.

3.  Black box on top of computer must have knob set to "B".
knob.jpg (9778 bytes)

4. After setting the switches above, turn on the power:
microscopeonoff.gif (12591 bytes)

4. Turn on the electronically controlled microscope stage.
prior.jpg (18695 bytes)

6. Turn on the cooled CCD camera.
camera-power.jpg (12369 bytes)

7. The computer is usually left on.  Check the screen is on and wiggle the mouse or hit the keyboard to see if computer is sleeping.
If the computer is not on, turn it on.

8. Run the software.
icon.gif (924 bytes)

 

Click Here for Function Keys.


To Change Objective Magnification

Please change the objective by using the point-and-click menu or by the keypad.
change_objective.gif (9082 bytes)
Before changing objective, make sure to lower the stage so that the lens may crash and destroy your sample!
focus.jpg (56789 bytes)
The button marked "ESC" drops the stage.  Pressing it again will bring the stage back to the same position.
dropstage.gif (41256 bytes)

 


What you see by eye and what you see using the camera

byeye.jpg (7486 bytes)  onscreen.jpg (10541 bytes)

What you see by eye through the microscope occulars is a mirror image of what you see on the computer screen.
Also, the image on the computer screen is only the center portion of the full field you can see by eye.

 


Nomarski / Differential Interference Contrast

There are four filters that must be adjusted properly for DIC microscopy.

  1. Polarizer on bottom directly above the field diaphagm.
    field_diaphragm.jpg (28741 bytes)
  2. Polarizer in body of microscope that slides in and out. [picture]
  3. 1/4 wave plate that slides in and out immediately above objective lenses. [picture]
  4. 1/4 wave plate in condenser, computer controlled.

First, the sample must be in focus and set for Koehler illumination with the field diaphragm at the edge of the field and the iris diaphragm (computer controlled) fully open.

The polarizers must be crossed with the two 1/4 wave plates out of the light path.

Then both 1/4 wave plates are reintroduced into the light path for DIC.

The iris diaphragm inside the condenser may be adjusted via computer for increased contrast.


Making Montages

Using the Make Montage Macros, field sizes are as follows:

Objective X (width) Y (height)
10X Plan N.A. 0.30 824 um 658 um
20X Plan N.A. 0.50 412 um 329 um
montagexy.jpg (26826 bytes)

 

 

Click Here for Function Keys.

Advanced technical info here.


return to index next pageemailIcon.gif (398 bytes)